PACIFIC
ELECTRONICS TONE VISUAL NURSE CALL SYSTEMS
FORMERLY AUTH-FLORENCE - NC-9100 STAFF STATION NS-2/NS-3 |
|
NS-2 / NS-3
Pacific Electronics NS2/3 Staff Stations are used in the
NC-9100 Nurse Call System to activate call signals from
lounges, examining
rooms, day rooms and other areas.
OPERATION
When the call push button on the staff station is pressed,
the following indicators light: The red Call Placed LED
on the station, the red Call In LED’s on the duty
stations, the dome and zone lamps, and the room and “Call
Placed” LED’s on the annunciator. At the same
time, audible tones are ouputted at the annunciator and
duty stations.
EQUIPMENT
The NS3 staff station has push buttons
for code blue, emergency, and normal call origination. It
also has LED indicators for Call Placed.
The NS2 has push buttons for emergency
and normal call origination. It also has LED indicators
for Call Placed.
It is furnished with a plastic faceplate which mounts to
a single gang back box.
SPECIFICATIONS
Front Plate: Plastic single gang plate
which mounts to back box.
Control (s): Code Blue push button to originate
code blue calls. Emergency Call push button to originate
emergency calls.
Staff Call push button to originate normal calls.
LED’s: Red for Call Placed on each switch.
Terminations: Six Position terminal block.
Mounting: Flush
Power Requirements: 24 VDC
|
| |